Remote commands exclusive to phase noise markers
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO.............................................. 745
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?........................................... 745
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe]........................................... 746
CALCulate<n>:MARKer<m>:FUNCtion:PNOise:RESult?................................................... 746
CALCulate<n>:MARKer<m>:FUNCtion:PNOise[:STATe]................................................... 747
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO <State>
This command turns an automatic peak search for the fixed reference marker at the
end of a sweep on and off.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:DELT:FUNC:PNO:AUTO ON
Activates an automatic peak search for the reference marker in
a phase-noise measurement.
Manual operation: See " Defining Reference Point " on page 445
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
This command queries the result of a phase noise measurement.
If necessary, the command activates the measurement first.
This command is only available in the Spectrum application.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<PhaseNoise> numeric value
The difference in level between the reference point and the
noise power density at the position of the specified delta marker.
Example:
CALC:DELT2:FUNC:PNO:RES?
Outputs the result of phase-noise measurement of the delta-
marker 2.
Usage: Query only
Manual operation: See " Phase Noise Measurement State " on page 444