CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<Power> Signal power over the marker bandwidth.
Example: Activate the band power marker:
CALC:MARK:FUNC:BPOW:STAT ON
Select the density mode for the result:
CALC:MARK:FUNC:BPOW:MODE DENS
Query the result:
CALC:MARK:FUNC:BPOW:RES?
Response:
20dBm/Hz
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:SPAN <Span>
This command defines the bandwidth around the marker position.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Span> Frequency. The maximum span depends on the marker position
and R&S FPL1000 model.
*RST: 5% of current span
Default unit: Hz
Example:
CALC:MARK:FUNC:BPOW:SPAN 2MHz
Measures the band power over 2 MHz around the marker.
Manual operation: See " Span " on page 450
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe] <State>
This command turns markers for band power measurements on and off.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Setting Basic Spectrum RF Measurement Parameters