User Manual

Remote Commands
R&S
®
FPL1000
754User Manual 1178.3370.02 ─ 03
Example:
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:TIME?
Outputs the time values of the temporary markers.
Usage: Query only
Manual operation: See " n dB down Delta Value " on page 447
8.8.8.13 Signal Count Marker
The following commands control the frequency counter.
CALCulate<n>:MARKer<m>:COUNt................................................................................754
CALCulate<n>:MARKer<m>:COUNt:FREQuency?............................................................755
CALCulate<n>:MARKer<m>:COUNt:RESolution...............................................................755
CALCulate<n>:MARKer<m>:COUNt <State>
This command turns the frequency counter at the marker position on and off.
The frequency counter works for one marker only. If you perform a frequency count
with another marker, the R&S FPL1000 deactivates the frequency count of the first
marker.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 536.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Setting Basic Spectrum RF Measurement Parameters