User Manual
Remote Commands
R&S
®
FPL1000
768User Manual 1178.3370.02 ─ 03
Example:
CALC:DLIN2:STAT ON
Turns on display line 2.
CALCulate<n>:FLINe<k> <Frequency>
This command defines the position of a frequency line.
Suffix:
<n>
.
Window
<k> Limit line
Parameters:
<Frequency> Note that you can not set a frequency line to a position that is
outside the current span.
Range: 0 Hz to Fmax
*RST: (STATe to OFF)
Example:
CALC:FLIN2 120MHz
Sets frequency line 2 to a frequency of 120 MHz.
Manual operation: See " Vertical Line 1 / Vertical Line 2 " on page 490
CALCulate<n>:FLINe<k>:STATe <State>
This command turns a frequency line on and off
Suffix:
<n>
.
Window
<k> Limit line
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:FLIN2:STAT ON
Turns frequency line 2 on.
CALCulate<n>:TLINe<k> <Time>
This command defines the position of a time line.
Suffix:
<n>
.
Window
<k> Limit line
Parameters:
<Time> Note that you can not set a time line to a position that is higher
than the current sweep time.
Range: 0 s to 1600 s
*RST: (STATe to OFF)
Setting Basic Spectrum RF Measurement Parameters