User Manual
Remote Commands
R&S
®
FPL1000
778User Manual 1178.3370.02 ─ 03
CALCulate<n>:LIMit<k>:UPPer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
Window
<k> Limit line
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See " Y-Axis " on page 500
CALCulate<n>:LIMit<k>:UPPer:STATe <State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<k>:NAME on page 775.
Suffix:
<n>
.
irrelevant
<k> Limit line
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Visibility " on page 497
CALCulate<n>:LIMit<k>:UPPer:THReshold <Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S FPL1000 uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
<k> Limit line
Parameters:
<Limit> Numeric value.
The unit depends on CALCulate<n>:LIMit<k>:UNIT
on page 775.
*RST: -200
Default unit: dBm
Manual operation: See " Threshold " on page 499
Setting Basic Spectrum RF Measurement Parameters