User Manual
Remote Commands
R&S
®
FPL1000
779User Manual 1178.3370.02 ─ 03
Managing Limit Lines
Useful commands for managing limit lines described in the R&S FPL1000 User Man-
ual:
●
MMEM:SEL[:ITEM]:LIN:ALL
●
MMEM:STOR:TYPE
Remote commands exclusive to managing limit lines:
CALCulate<n>:LIMit<k>:ACTive?.................................................................................... 779
CALCulate<n>:LIMit<k>:COPY........................................................................................779
CALCulate<n>:LIMit<k>:DELete......................................................................................780
CALCulate<n>:LIMit<k>:STATe.......................................................................................780
CALCulate<n>:LIMit<k>:TRACe<t>:CHECk......................................................................780
MMEMory:LOAD<n>:LIMit.............................................................................................. 781
MMEMory:STORe<n>:LIMit............................................................................................ 781
CALCulate<n>:LIMit<k>:ACTive?
This command queries the names of all active limit lines.
Suffix:
<n>, <k>
.
irrelevant
Return values:
<LimitLines> String containing the names of all active limit lines in alphabeti-
cal order.
Example:
CALC:LIM:ACT?
Queries the names of all active limit lines.
Usage: Query only
Manual operation: See " Visibility " on page 497
CALCulate<n>:LIMit<k>:COPY <Line>
This command copies a limit line.
Suffix:
<n>
.
Window
<k> Limit line
Parameters:
<Line> 1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Example:
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.