MMEMory:SELect:CHANnel[:ITEM]:ALL
MMEMory:SELect[:ITEM]:ALL
This command includes all items when storing or loading a configuration file.
The items are:
●
Hardware configuration: MMEMory:SELect[:ITEM]:HWSettings
●
Limit lines: MMEMory:SELect[:ITEM]:LINes:ALL
●
Spectrogram data: MMEMory:SELect[:ITEM]:SGRam
●
Trace data: MMEMory:SELect[:ITEM]:TRACe[:ACTive]
●
Transducers: MMEMory:SELect[:ITEM]:TRANsducer:ALL
Example:
MMEM:SEL:ALL
Usage: Event
Manual operation: See " Items: " on page 105
MMEMory:SELect:CHANnel[:ITEM]:DEFault
MMEMory:SELect[:ITEM]:DEFault
This command selects the current settings as the only item to store to and load from a
configuration file.
Usage:
Event
Manual operation: See " Items: " on page 105
MMEMory:SELect:CHANnel[:ITEM]:HWSettings <State>
MMEMory:SELect[:ITEM]:HWSettings <State>
This command includes or excludes measurement (hardware) settings when storing or
loading a configuration file.
Measurement settings include:
●
general channel configuration
●
measurement hardware configuration including markers
●
limit lines
Note that a configuration may include no more than 8 limit lines. This number
includes active limit lines as well as inactive limit lines that were used last.
Therefore the combination of inactivate limit lines depends on the sequence of use
with MMEMory:LOAD:STATe.
●
color settings
●
configuration for the hardcopy output
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
MMEM:SEL:HWS ON
Managing Settings and Results