Usage: Query only
Manual operation: See " Alignment Results: " on page 160
DIAGnostic:SERVice:INPut:PULSed:CFRequency <Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect].
Parameters:
<Frequency> Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S FPL1000
uses the next available frequency. Example: a frequency of
20 MHz is rounded up to the next available frequency (25 MHz).
*RST: 50 MHz
Default unit: Hz
Manual operation: See " Calibration Frequency RF " on page 145
DIAGnostic:SERVice:INPut:RF[:SPECtrum] <Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth> NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
DIAGnostic:SERVice:INPut[:SELect] <Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S FPL1000.
Parameters:
<Signal> CALibration
Uses the calibration signal as RF input.
RF
Uses the signal from the RF input.
*RST: RF
Example:
DIAG:SERV:INP CAL
Uses the calibration signal as RF input.
Manual operation: See " NONE " on page 145
See " Calibration Frequency RF " on page 145