User Manual

Remote Commands
R&S
®
FPL1000
819User Manual 1178.3370.02 ─ 03
DIAGnostic:SERVice:STESt:RESult?
This command queries the self-test results.
Return values:
<Results> String of data containing the results.
The rows of the self-test result table are separated by commas.
Example:
DIAG:SERV:STES:RES?
would return, e.g.
"Total Selftest Status:
PASSED", "Date (dd/mm/yyyy): 09/07/2004 TIME:
16:24:54", "Runtime: 00:06", "...
Usage: Query only
SOURce:TEMPerature:FRONtend?
This command queries the current frontend temperature of the R&S FPL1000.
During self-alignment, the instrument's (frontend) temperature is also measured (as
soon as the instrument has warmed up completely). This temperature is used as a ref-
erence for a continuous temperature check during operation. If the current temperature
deviates from the stored self-alignment temperature by a certain degree, a warning is
displayed in the status bar indicating the resulting deviation in the measured power lev-
els. A status bit in the STATUs:QUEStionable:TEMPerature register indicates a
possible deviation.
Return values:
<Temperature> Temperature in degrees Celsius.
Example:
SOUR:TEMP:FRON?
Queries the temperature of the frontend sensor.
Usage: Query only
8.10.3 Working with Transducers
The following commands configure and control transducer factors.
Useful commands for transducer management described elsewhere
MMEMory:SELect[:ITEM]:TRANsducer:ALL on page 795
Remote commands exclusive to transducer management
[SENSe:]CORRection:TRANsducer:ADJust:RLEVel[:STATe]............................................. 820
[SENSe:]CORRection:TRANsducer:CATalog?..................................................................820
[SENSe:]CORRection:TRANsducer:COMMent..................................................................820
[SENSe:]CORRection:TRANsducer:DATA........................................................................821
[SENSe:]CORRection:TRANsducer:DELete......................................................................821
[SENSe:]CORRection:TRANsducer:SCALing....................................................................821
[SENSe:]CORRection:TRANsducer:SELect......................................................................822
Configuring the R&S FPL1000