User Manual
Remote Commands
R&S
®
FPL1000
821User Manual 1178.3370.02 ─ 03
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Comment> *RST: (empty comment)
Manual operation: See " Comment " on page 152
[SENSe:]CORRection:TRANsducer:DATA
<Frequency>,<Level>[,<Frequency>,<Level>]
This command configures transducer factors for specific trace points. A set of trans-
ducer factors defines an interpolated transducer line and can be stored on the instru-
ment.
Parameters:
<Frequency> The unit for <Frequency> is Hz, which may or may not be omit-
ted. Frequencies have to be sorted in ascending order.
<Level> The unit for <Level> depends on[SENSe:]CORRection:
TRANsducer:UNIT .
Example:
SENSe1:CORRection:TRANsducer:UNIT 'DB'
// Frequency Span 0 Hz to 4 Ghz
SENSe1:CORRection:TRANsducer:DATA 0,8,2e6,5,4GHz,3
Creates the transducer points:
Manual operation: See " Data Points " on page 153
Frequency Level
0 Hz 8 dB
2 GHz 5 dB
4 GHz 3 dB
[SENSe:]CORRection:TRANsducer:DELete
This command deletes the currently selected transducer factor.
Before you can use the command, you have to select a transducer.
Example:
CORR:TRAN:DEL
Usage: Event
Manual operation: See " Delete Line " on page 151
[SENSe:]CORRection:TRANsducer:SCALing <ScalingType>
This command selects the frequency scaling of the transducer factor.
Parameters:
<ScalingType> LINear | LOGarithmic
*RST: LINear