832User Manual 1178.3370.02 ─ 03
SYSTem:DISPlay:UPDate <State>
This command turns the display during remote operation on and off.
If on, the R&S FPL1000 updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:DISP:UPD ON
Manual operation: See "Remote Display Update" on page 203
SYSTem:ERRor:DISPlay
<State>
This command the error display during remote operation on and off.
If activated, the R&S FPL1000 displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:ERR:DISP ON
Manual operation: See "Display Remote Errors" on page 203
SYSTem:IDENtify:FACTory
This command resets the query to *IDN? to its default value.
Usage:
Event
Manual operation: See "Reset to Factory String" on page 202
SYSTem:IDENtify[:STRing] <String>
This command defines the response to *IDN?.
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 202
SYSTem:KLOCk <State>
This command activates the local lockout (remote control) or returns to the local mode.