User Manual

Remote Commands
R&S
®
FPL1000
837User Manual 1178.3370.02 ─ 03
SYSTem:ERRor:EXTended? <MessageType>[, <ChannelName>]
This command queries all system messages, or all messages of a defined type, dis-
played in the status bar for a specific channel setup (application).
Note: This command queries the strings displayed for manual operation. For remote
programs, do not define processing steps depending on these results. Instead, query
the results of the STATus:QUEStionable:EXTended:INFO status register, which
indicates whether messages of a certain type have occurred (see "STATus:QUEStion-
able:EXTended:INFO Register" on page 191).
Query parameters:
<MessageType> ALL | INFO | WARNing | ERRor | FATal | MESSage
<ChannelName> String containing the name of the channel setup.
The parameter is optional. If you omit it, the command works for
the currently active channel setup.
Return values:
<Messages> String containing all messages of the selected type for the speci-
fied channel setup. Each message is separated by a comma
and inserted in parentheses. If no messages are available,
empty parentheses are returned.
Example:
SYST:ERR:EXT? ALL
Returns all messages for the currently active application, e.g.
"Message 1","Message 2".
Example:
SYST:ERR:EXT? FAT,'Spectrum2'
Queries fatal errors in the 'Spectrum2' application. If none have
occurred, the result is: "
".
Usage: Query only
SYSTem:ERRor:LIST? [<MessType>]
This command queries the error messages that occur during R&S FPL1000 operation.
Query parameters:
<MessType> SMSG
(default) Queries the system messages which occurred during
manual operation.
REMote
Queries the error messages that occurred during remote opera-
tion.
Note: The remote error list is automatically cleared when the
R&S FPL1000 is shut down.
Return values:
<SystemMessages> String containing all messages in the "System Messages" table.
Configuring the R&S FPL1000