Applications, Measurement Channels, and Result Displays
R&S
®
FPL1000
85User Manual 1178.3370.02 ─ 03
New Channel Setup ← Defining a channel setup
The application selected on this tab of the dialog box activates a new channel setup,
i.e. a new tab in the display.
Note: The channel setups are labeled with the application name. If that name already
exists, a sequential number is added. You can change the name of the channel setup
by double-tapping the name in the channel setup bar and entering a new name.
For an overview of default names see Table 8-2.
Remote command:
INSTrument:CREate[:NEW] on page 527
INSTrument[:SELect] on page 529
Replace Current Channel Setup ← Defining a channel setup
The application selected on this tab of the dialog box is started in the currently dis-
played channel setup, replacing the current measurement.
Remote command:
INSTrument:CREate:REPLace on page 527
Duplicate Current Channel ← Defining a channel setup
The currently active channel setup can be duplicated, i.e. a new channel setup of the
same type and with the identical measurement settings is started. The name of the
new channel setup is the same as the copied channel setup, extended by a consecu-
tive number (e.g. "Spectrum" -> "Spectrum 2" ).
Remote command:
INSTrument:CREate:DUPLicate on page 527
Closing a channel setup
To close a channel setup, simply close the corresponding tab by selecting the "x" next
to the channel setup name.
Remote command:
INSTrument:DELete on page 528
5.4 Running a Sequence of Measurements
Only one measurement can be performed at any time, namely the one in the currently
active channel setup. However, in order to perform the configured measurements con-
secutively, a Sequencer function is provided.
● The Sequencer Concept......................................................................................... 85
● Sequencer Settings.................................................................................................88
● How to Set Up the Sequencer.................................................................................89
5.4.1 The Sequencer Concept
The instrument can only activate one specific channel setup at any time. Thus, only
one measurement can be performed at any time, namely the one in the currently active
channel setup. However, in order to perform the configured measurements consecu-
Running a Sequence of Measurements