2. In the Identify column, activate the function Start (starting with Device No. 1).
The LEDs on the associated amplifier flash. This identifies the amplifier with the (tem-
porary) order number 1.
3. Under Device, change the order number according to the position of the identified am-
plifier in the rack.
4. Repeat steps 2 to 3 for the other amplifiers in the rack.
The procedure is completed when each order number in the Device column correctly
corresponds to the position of the associated amplifier in the rack.
3.3 Clearing Event Memory
There are four event memories for each of the R&S NetCCU800, exciter and OS (output
stage) modules.
Summary
Status
Warning
Fault
These event memories need to be cleared for all the modules before the transmitter begins
regular transmission.
1. Select NetCCU > Logbook > Status.
The Status window opens.
2. Select the Clear Logbook command in the context menu.
The entries are cleared.
3. Repeat the procedure for each of the event memories listed.