RT-Zxx_High_voltage_dat-sw_en_5214-2362-22_v1600_Cover.indd 1 Data Sheet | 16.00 Test & Measurement R&S®RT-Zxx High-Voltage and Current Probes Specifications 09.11.
Version 16.00, November 2016 CONTENTS Definitions.............................................................................................................................................................................................. 3 Probe/oscilloscope chart ........................................................................................................................................................................ 4 R&S®RT-ZH03 high-voltage passive probe ..................................
Version 16.00, November 2016 Definitions General Product data applies under the following conditions: Three hours storage at ambient temperature followed by 30 minutes warm-up operation Specified environmental conditions met Recommended calibration interval adhered to Specifications with limits Represent warranted product performance by means of a range of values for the specified parameter.
Version 16.
Version 16.
Version 16.00, November 2016 R&S®RT-ZH03 high-voltage passive probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 16.00, November 2016 R&S®RT-ZH03 maximum rated sine-wave root mean square voltage versus frequency.
Version 16.00, November 2016 General data Temperature Temperature loading Climatic loading Altitude Safety Mechanical data Dimensions Weight Probe interface Connector 8 operating temperature range operation 0 °C to +40 °C 80 % relative humidity without condensation up to 2000 m in line with Low Voltage Directive 2006/95/EC, IEC/EN 61010-31 (pollution degree 2) diameter of probe tip cable length probe only approx. 5 mm (0.2 in) approx. 1.3 m (51 in) approx. 55 g (0.
Version 16.00, November 2016 R&S®RT-ZH10/-ZH11 high-voltage passive probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 16.00, November 2016 R&S®RT-ZH10/-ZH11 maximum rated sine-wave root mean square voltage versus frequency, CAT I. R&S®RT-ZH10/-ZH11 maximum root mean square voltage versus peak pulse voltage, CAT I. R&S®RT-ZH10/-ZH11 maximum pulse derating, CAT I.
Version 16.
Version 16.00, November 2016 R&S®RT-ZI10(C)/-ZI11 isolated probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. The R&S RT-ZI10/-ZI11 must be used only with insulated oscilloscopes provided with touch-protected inputs. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 16.00, November 2016 R&S®RT-ZI10C Step response Rise time Frequency response Bandwidth Input impedance DC input resistance Input capacitance DC characteristics Attenuation Base unit Use with Input capacitance Input coupling system, 10 % to 90 % 700 ps (meas.) system, –3 dB, starting at DC > 500 MHz (meas.) system system 10 MΩ ± 1 % (meas.) 11 pF (meas.
Version 16.00, November 2016 R&S®RT-ZI10(C)/-ZI11 maximum rated sine-wave root mean square voltage between probe tip and probe reference terminal versus frequency (CAT III).
Version 16.
Version 16.00, November 2016 R&S®RT-ZD01 high-voltage differential probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 16.00, November 2016 Maximum rated input voltage Continuous voltage derated, see figure, each signal socket to ground 1000 V (RMS), CAT III Base unit Input coupling 1 MΩ AC/DC Sine-wave RMS voltage [V] 10000 1000 100 10 100E3 1E6 10E6 100E6 Frequency [Hz] Maximum rated sine-wave root mean square voltage versus frequency.
Version 16.
Version 16.00, November 2016 R&S®RT-ZD02/-ZD08 high-voltage differential probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details. Step response Rise time Frequency response Bandwidth Common mode rejection Input impedance DC input resistance Input capacitance DC characteristics Max.
Version 16.
Version 16.00, November 2016 R&S®RT-ZC02/-ZC03 current probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details. Sensitivity setting Step response Rise time Frequency response Bandwidth DC characteristics Dynamic range Sensitivity error Temperature drift, sensitivity Zero error R&S®RT-ZC02 0.01 V/A 10 % to 90 % 5 µs (meas.
Version 16.00, November 2016 R&S®RT-ZC03 Step response Rise time Frequency response Bandwidth DC characteristics Dynamic range Sensitivity Sensitivity error Temperature drift, sensitivity Zero error 10 % to 90 % 1 µs (meas.) –3 dB, starting at DC 100 kHz (meas.) +23 °C ±1 °C ±30 A 0.1 V/A ±1 % (meas.) ± 0.01 %/°C (meas.) ±2 mA (meas.
Version 16.
Version 16.00, November 2016 R&S®RT-ZC05B/-ZC10(B)/-ZC15B/-ZC20(B)/-ZC30 current probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 16.00, November 2016 Step response Rise time Frequency response Bandwidth Input impedance DC characteristics Sensitivity Sensitivity error Zero error AC characteristics AC sensitivity error (sinusoidal, 45 Hz to 66 Hz) Measurement due to external magnetic fields Maximum rated input Maximum continuous current Maximum transient current Other Noise R&S®RT-ZC15B R&S®RT-ZC20(B) 10 % to 90 % 7 ns (meas.) 3.5 ns (meas.) –3 dB, starting at DC 50 MHz (meas.) see figure on page 28 100 MHz (meas.
Version 16.00, November 2016 R&S®RT-ZC30 Step response Rise time Frequency response Bandwidth Input impedance DC characteristics Sensitivity Sensitivity error Zero error AC characteristics AC sensitivity error (sinusoidal, 45 Hz to 66 Hz) Measurement due to external magnetic fields Maximum rated input Maximum continuous current Maximum transient current Other Noise 26 10 % to 90 % 2.9 ns (meas.) –3 dB, starting at DC 120 MHz (meas.
Version 16.00, November 2016 General data R&S®RT-ZC05B/ R&S®RT-ZC10(B) Temperature Temperature loading Climatic loading Altitude Safety operating temperature range storage temperature range operation EMC Calibration interval Mechanical data Dimensions Weight Probe interface Connector Supply voltage max.
Version 16.00, November 2016 Maximum rated sine-wave root mean square input current versus frequency. Input impedance (meas.). Maximum rated sine-wave root mean square input current versus frequency. Input impedance (meas.).
Version 16.00, November 2016 R&S®RT-ZA13 probe power supply Electrical data Number of channels Output voltage Maximum output current Power requirements Maximum rated input power sum total of all channels 4 ±12 V ± 0.5 V 2.5 A 100 V to 240 V, 50/60 Hz 170 W General data Safety EMC Mechanical data Dimensions Weight Connector in line with EN 61010-1 in line with EN 61326-1 (class B equipment), EN 61000-3-2, EN 61000-3-3 W×H×L approx. 80 mm × 119 mm × 200 mm (3.1 in × 4.7 in × 7.9 in) approx. 1.1 kg (2.
Version 16.00, November 2016 Ordering information Designation High-voltage passive probes 250 MHz High-Voltage Probe, passive, 100:1, 100 MΩ, 6.5 pF, 850 V (RMS) Incl. adjustment tool; coding clips (set) 2 × 4 colors; signal pin (2); sprung hook 5 mm; ground lead 14 cm; insulating cap; protective cap; operating manual 400 MHz High-Voltage Probe, passive, 100:1, 50 MΩ, 7.5 pF, 1 kV (RMS) Incl. adjustment tool; BNC adapter 5.0-L; coding rings (set) 3 × 4 colors; flexible adapter 5.
Version 16.00, November 2016 Designation Differential probes 100 MHz Differential Probe, ±1.4 kV, 1 kV RMS (CAT III), BNC Incl. sprung hook 4 mm (2); USB power cord; carrying case; operating manual 200 MHz Differential Probe, ±20 V, BNC Incl. safety alligator clip 4 mm (2); sprung hook 4 mm (2); USB power cord; 9 V battery; carrying case; operating manual 800 MHz Differential Probe, ±15 V, BNC Incl.
Version 16.00, November 2016 Designation Accessories and sets Mini Clips, contains: mini clip (10) Micro Clips, contains: micro clip (4) Lead Set, contains: lead 6 cm (2.4 in) (5); lead 15 cm (5.9 in) (5) Probe Box to N/USB Adapter SMA(f) to BNC(m) Adapter Probe Power Supply 32 Rohde & Schwarz R&S®RT-Zxx High-Voltage and Current Probes Type Order No. R&S®RT-ZA4 R&S®RT-ZA5 R&S®RT-ZA6 R&S®RT-ZA9 R&S®RT-ZA10 R&S®RT-ZA13 1416.0428.02 1416.0434.02 1416.0440.02 1417.0909.02 1416.0457.02 1409.7789.
Version 16.
Version 16.
Version 16.
Sustainable product design ❙❙ Environmental compatibility and eco-footprint ❙❙ Energy efficiency and low emissions Service that adds value Worldwide Local and personalized ❙ Customized and flexible ❙ Uncompromising quality ❙ Long-term dependability ❙❙ Longevity and optimized total cost of ownership ❙ ❙ Certified Quality Management ISO 9001 Certified Environmental Management ISO 14001 Rohde & Schwarz GmbH & Co. KG www.rohde-schwarz.