Brochure
Instantaneous power consumption
Voltage at DUT
Current consumption of DUT
Zoom in µA range
12mA
30µA
V
I
V
I
Rohde & Schwarz Probes and accessories for Rohde & Schwarz oscilloscopes 23
Measure up to eight power rails at the same time
with highest accuracy
One R&S®RTE, R&S®RTO or R&S®RTP oscilloscope sup-
ports up to two R&S®RT-ZVC current probes so that it is
possible to observe eight power domains in parallel with
a DC accuracy of 0.1 % for voltage measurements and
0.2 % for current measurements. Ramp-up processes and
power rail tolerances can easily be tested with this probe.
The oscilloscope's SCPI remote control enables automatic
testing.
The R&S®RT-ZVC probe provides an extraordinarily high dynamic range for measuring both active state currents and sleep currents, in this example
12mA and 30µA. Automated measurements make it possible to calculate the total energy consumption.
Digital acquisition system
The R&S®RT-ZVC probe's digital acquisition system provides 18-bit resolution, a
5Msample/s sampling rate and 1MHz bandwidth. Each voltage and current input pair
forms a high dynamic range power measurement system.
High bandwidth with flexible filtering for noise
reduction
A bandwidth of 1 MHz and a sampling rate of
5Msample/s allows you to capture fast current pulses. To
analyze the overall power consumption of battery powered
devices, the very low sleep mode currents have to be cap-
tured at the same time. For very high dynamic range mea-
surements, the integrated lowpass filter reduces the band-
width down to 5 kHz and minimizes overall system noise.
Current ranges
Low-gain mode, shunt
❙ ±4.5µA; ±45µA, 10 kΩ
❙ ±4.5mA; ±45mA, 10 Ω
❙ ±4.5A; ±10A, 10 mΩ
❙ ±45mV
1)
; ±450mV
1)
, external
Voltage ranges
❙ ±1.88V
❙ ±3.75V
❙ ±7.5V
❙ ±15V
1)
Current range depends on shunt value.
Digital backend (4 in total)
A/D
18 bit
5 Msample/s
A/D
18 bit
5 Msample/s
5 Msample/s
Trigger
Analog frontend
Deskew
Lowpass
Probes_and_accessories_bro_en_3606-8866-12_v1400.indd 23 23.07.2019 13:48:01