Brochure

Rohde & Schwarz Probes and accessories for Rohde & Schwarz oscilloscopes 29
Model Bandwidth Attenuation
factor
Input
impedance
Dynamic
range
Comment Order No.
Passive
R&S®RT-ZH03 250 MHz 100:1 100 MΩ || 6.5 pF 850 V (RMS) robust 5mm probe tip 1333.0873.02
R&S®RT-ZH10 400 MHz 100:1 50 MΩ || 7.5 pF 1000 V (RMS),
6000 V (peak)
1000 V (RMS) CAT II, 5mm probe
tip, spring-loaded
1409.7720.02
R&S®RT-ZH11 400 MHz 1000:1 50 MΩ || 7.5 pF 1000 V (RMS),
6000 V (peak)
1000 V (RMS) CAT II, 5mm probe
tip, spring-loaded
1409.7737.02
R&S®RT-ZI11 500 MHz 100:1 100 MΩ || 4.6 pF 1000 V (RMS) 600 V (RMS) CAT IV,
1000V(RMS) CAT III,
3540 V (RMS) CAT 0, for
R&S®Scope Rider RTH only
1326.1810.02
Active, differential
R&S®RT-ZD002 25 MHz 10:1/100:1 8 MΩ || 2.75 pF ±700 V 1000 V (RMS) CAT III 1337.9700.02
R&S®RT-ZD003 25 MHz 20:1/200:1 8 MΩ || 2.75 pF ±1400 V 1000 V (RMS) CAT III 1337.9800.02
R&S®RT-ZD01 100MHz 100:1/1000:1 8 MΩ || 3.5 pF ±1400 V 1000 V (RMS) CAT III 1422.0703.02
R&S®RT-ZHD07 200 MHz 25:1/250:1 5 MΩ || 2.5 pF ±750 V 300 V (RMS) CAT III 1800.2307.02
R&S®RT-ZHD15 100 MHz 50:1/500:1 10 MΩ || 2 pF ±1500 V 1000 V (RMS) CAT III 1800.2107.02
R&S®RT-ZHD16 200 MHz 50:1/500:1 10 MΩ || 2 pF ±1500 V 1000 V (RMS) CAT III 1800.2207.02
R&S®RT-ZHD60 100 MHz 100:1/1000:1 40 MΩ || 2 pF ±6000 V 1000 V (RMS) CAT III 1800.2007.02
Accessory
R&S®RT-ZA24 replacement kit for R&S®RT-ZHD
probes
1800.2707.00
Measuring voltages up to 1000V (RMS) with
500MHz bandwidth
Measuring in high voltage environments requires special
safety precautions in line with the European Low Volt-
age Directive that depend on the actual measurement
environment.
With a 500MHz bandwidth, the R&S®RT-ZI11 passive probes
can be used to measure fast switching edges of GaN power
electronics designs. Designed for the R&S®Scope Rider
oscilloscope with its isolated channels, the probes are ideal
for floating measurements of voltages with common mode
voltages upto1000V(RMS).
The R&S®Scope RiderRTH lets you safely measure voltag-
es up to 1000V in CATIII environments or 600V in CATIV
environments. In combination with the R&S®RT-ZI11 100:1
passive high voltage probe, a bandwidth of 500MHz is
achieved. This makes the solution interesting for measure-
ments on GaN based power electronics components.
Probes_and_accessories_bro_en_3606-8866-12_v1400.indd 29 23.07.2019 13:48:13