Datasheet

Version 20.00, November 2019
Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes 19
R&S
®
RT-ZD01 high voltage differential probe
All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
R&S
®
RT-ZD01
Attenuation setting
100:1
1000:1
Step response
Rise time
10 % to 90 %
< 3.5 ns (meas.)
Frequency response
Bandwidth
starting at DC, calculated from 0.35/rise time
100 MHz
Common mode rejection
DC to 100 Hz
80 dB (meas.)
100 Hz to 1 MHz
50 dB (meas.)
Input impedance
DC input resistance
differential (between signal sockets)
8 MΩ
single-ended (each signal socket to ground)
4 MΩ
Input capacitance
differential (between signal sockets)
3.5 pF (meas.)
single-ended (each signal socket to ground)
7 pF (meas.)
DC characteristics
Attenuation error
±2 %
Zero error
referenced to probe input
±0.5 V (meas.)
±5 V (meas.)
Max. differential input
between signal sockets
±140 V
±1400 V
Operating voltage window
each signal socket to ground
±1400 V
Noise voltage
referenced to probe input
90 mV (RMS)
(meas.)
0.9 V (RMS)
(meas.)