Datasheet

Version 20.00, November 2019
22 Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes
R&S
®
RT-ZD02/-ZD08 high voltage differential probes
All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 50 Ω.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
R&S
®
RT-ZD02
R&S
®
RT-ZD08
Step response
Rise time
10 % to 90 %
1.75 ns (meas.)
437 ps (meas.)
Frequency response
Bandwidth
3 dB, starting at DC,
calculated from 0.35/rise time
200 MHz
800 MHz
Common mode rejection
DC to 100 Hz
80 dB (meas.)
60 dB (meas.)
100 Hz to 10 MHz
50 dB (meas.)
100 Hz to 500 MHz
15 dB (meas.)
Input impedance
DC input resistance
differential (between signal sockets)
1 (meas.)
200 kΩ (meas.)
single-ended (each signal socket to ground)
500 kΩ (meas.)
100 kΩ (meas.)
Input capacitance
differential (between signal sockets)
3.5 pF (meas.)
1 pF (meas.)
single-ended (each signal socket to ground)
7 pF (meas.)
2 pF (meas.)
DC characteristics
Max. differential input
between signal sockets
±20 V
±15 V
Operating voltage window
each signal socket to ground
±60 V
±30 V
Attenuation
10:1
10:1
Attenuation error
probe only, with ideal 50 Ω load impedance
±1 % (meas.)
±2 % (meas.)
Zero error
at probe output
±2 mV (meas.)
±5 mV (meas.)
Maximum rated input voltage
DC peak voltage
single-ended (each signal socket to ground)
±60 V
±40 V
AC peak voltage
single-ended (each signal socket to ground)
±60 V
±40 V
Base unit
Input coupling
50 Ω