Datasheet

Version 20.00, November 2019
24 Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes
R&S
®
RT-ZHD07 high voltage differential probe
All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Attenuation setting
25:1
250:1
Step response
Rise time
10 % to 90 %, both attenuations
< 2 ns
Frequency response
Bandwidth
starting at DC, calculated from 0.4/rise time
200 MHz
Common mode rejection
DC to 60 Hz
+15 °C to +35 °C
> 80 dB
0 °C to +50 °C
> 75 dB
60 Hz to 1 kHz
70 dB (meas.)
65 dB (meas.)
1 kHz to 1 MHz
55 dB (meas.)
55 dB (meas.)
1 MHz to 50 MHz
35 dB (meas.)
20 dB (meas.)
Input impedance
DC input resistance
differential (between signal sockets)
5
single-ended (each signal socket to ground)
2.5
Input capacitance
differential (between signal sockets)
2.5 pF (meas.)
single-ended (each signal socket to ground)
5 pF (meas.)
DC characteristics
Attenuation error
after applying digital correction factors
±0.5 %
Temperature drift, attenuation
±60 ppm/°C
Zero error
after applying digital correction factors,
referenced to probe input
+15 °C to +35 °C
±12.5 mV
±35 mV
0 °C to +50 °C
±25 mV
±55 mV
Temperature drift, zero error
referenced to probe input
±0.75 mV/°C
±1.12 mV/°C
referenced to probe output
±30 µV/°C
±4.5 µV/°C