Datasheet

Version 20.00, November 2019
Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes 25
Dynamic range
Differential input
between signal sockets
±75 V
±750 V
Offset compensation range
in both attenuations
±1000 V
Offset compensation error
offset compensation setting = 0 V
no additional error
offset compensation setting 0 V
±0.2 % of setting ±40 mV (meas.)
Operating voltage window
each signal socket to ground
±750 V
Noise voltage
referenced to probe input
12 mV (RMS)
(meas.)
40 mV (RMS)
(meas.)
Maximum rated input voltage
Continuous voltage
derated, see figure, each signal socket to ground
300 V (RMS), CAT III
600 V (RMS), CAT II
600 V (RMS)
Transient voltage
each signal socket to ground
±4500 V (peak)
Base unit
Input coupling
1 MΩ AC/DC