Datasheet
Version 20.00, November 2019
Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes 29
R&S
®
RT-ZHD15/-ZHD16 high voltage differential probe
All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Attenuation setting
50:1
500:1
Step response
Rise time
10 % to 90 %, both attenuations
R&S
®
RT-ZHD15
< 4 ns
R&S
®
RT-ZHD16
< 2 ns
Frequency response
Bandwidth
starting at DC, calculated from 0.4/rise time
R&S
®
RT-ZHD15
100 MHz
R&S
®
RT-ZHD16
200 MHz
Common mode rejection
DC to 60 Hz
+15 °C to +35 °C
> 80 dB
0 °C to +50 °C
> 75 dB
60 Hz to 1 kHz
70 dB (meas.)
65 dB (meas.)
1 kHz to 1 MHz
55 dB (meas.)
55 dB (meas.)
1 MHz to 50 MHz
35 dB (meas.)
20 dB (meas.)
Input impedance
DC input resistance
differential (between signal sockets)
10 MΩ
single-ended (each signal socket to ground)
5 MΩ
Input capacitance
differential (between signal sockets)
2 pF (meas.)
single-ended (each signal socket to ground)
4 pF (meas.)
DC characteristics
Attenuation error
after applying digital correction factors
±0.5 %
Temperature drift, attenuation
±60 ppm/°C