Datasheet
Version 20.00, November 2019
30 Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes
Zero error
after applying digital correction factors,
referenced to probe input
+15 °C to +35°C
±25 mV
±65 mV
0 °C to +50°C
±50 mV
±95 mV
Temperature drift, zero error
referenced to probe input
±1.5 mV/°C
±1.75 mV/°C
referenced to probe output
±30 µV/°C
±3.5 µV/°C
Dynamic range
Differential input
between signal sockets
±150 V
±1500 V
Offset compensation range
in both attenuations
±2000 V
Offset compensation error
offset compensation setting = 0 V
no additional error
offset compensation setting ≠ 0 V
±0.2 % of setting ± 80 mV (meas.)
Operating voltage window
each signal socket to ground
±1500 V
Noise voltage
referenced to probe input, (RMS)
R&S
®
RT-ZHD15
20 mV (meas.)
70 mV (meas.)
R&S
®
RT-ZHD16
25 mV (meas.)
80 mV (meas.)
Maximum rated input voltage
Continuous voltage
derated, see figure, each signal socket to ground
1000 V (RMS), CAT III
1000 V (RMS)
Transient voltage
each signal socket to ground
±6800 V (peak)
Base unit
Input coupling
1 MΩ AC/DC