Datasheet

Version 20.00, November 2019
Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes 31
Maximum rated sine-wave root mean square voltage versus frequency; each signal socket to ground
R&S
®
ProbeMeter
Specifications for measurement error apply only when offset compensation setting is 0 V. The R&S
®
ProbeMeter can be used to measure
differential and common mode voltages.
Measurement error,
differential mode and common mode
+15 °C to +35 °C
1000 V
±0.1 % of reading ±0.03 V
> 1000 V
±0.1 % of reading ±0.03 V (meas.)
0 °C to +50 °C
1000 V
±0.2 % of reading ±0.06 V
> 1000 V
±0.2 % of reading ±0.06 V (meas.)
Temperature drift
±60 ppm/°C of reading ±1.5 mV/°C
Common mode rejection,
for differential measurement
+15 °C to +35 °C
> 80 dB
0 °C to +50 °C
> 75 dB