Datasheet
Version 20.00, November 2019
Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes 33
R&S
®
RT-ZHD60 high voltage differential probe
All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Attenuation setting
100:1
1000:1
Step response
Rise time
10 % to 90 %, both attenuations
< 4 ns
Frequency response
Bandwidth
starting at DC, calculated from 0.4/rise time
100 MHz
Common mode rejection
DC to 60 Hz
+15 °C to +35 °C
> 80 dB
0 °C to +50 °C
> 75 dB
60 Hz to 1 kHz
70 dB (meas.)
65 dB (meas.)
1 kHz to 1 MHz
55 dB (meas.)
55 dB (meas.)
1 MHz to 50 MHz
35 dB (meas.)
20 dB (meas.)
Input impedance
DC input resistance
differential (between signal sockets)
40 MΩ
single-ended (each signal socket to ground)
20 MΩ
Input capacitance
differential (between signal sockets)
2 pF (meas.)
single-ended (each signal socket to ground)
4 pF (meas.)
DC characteristics
Attenuation error
after applying digital correction factors
±0.5 %
Temperature drift, attenuation
±80 ppm/°C
Zero error
after applying digital correction factors,
referenced to probe input
+15 °C to +35°C
±70 mV
±150 mV
0 °C to +50°C
±150 mV
±230 mV
Temperature drift, zero error
referenced to probe input
±5 mV/°C
±5.5 mV/°C
referenced to probe output
±50 µV/°C
±5.5 µV/°C