Datasheet

Version 20.00, November 2019
38 Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes
General data
R&S
®
RT-ZC02
R&S
®
RT-ZC03
Temperature
Temperature loading
operating temperature range
0 °C to +50 °C
storage temperature range,
with battery removed
20 °C to +85 °C
Climatic loading
80 % relative humidity for temperatures up to +31 °C,
decreasing linearly to 40 % at +50 °C
Altitude
operation
up to 2000 m
Safety
in line with EN 61010-1
in line with EN 61010-2-032 (pollution degree 2)
RoHS
in line with EN50581
EMC
in line with EN 61326-2-2
Calibration interval
2 years
Mechanical data
Dimensions
diameter of probe tip
approx. 32 mm (1.3 in)
approx. 25 mm (1.0 in)
cable length
approx. 2.0 m (79 in)
Weight
probe only
approx. 320 g (0.7 lb)
Probe interface
Connector
BNC
Battery type
9 V Alkaline battery, PP3,
MN 1604 or IEC6LR61
Battery lifetime
50 h (meas.)
25 h (meas.)