Datasheet
Version 20.00, November 2019
6 Rohde & Schwarz R&S
®
RT-Zxx High Voltage and Current Probes
R&S
®
RT-ZH03 high voltage passive probe
All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
R&S
®
RT-ZH03
Step response
Rise time
system, 10 % to 90 %
1.4 ns (meas.)
Frequency response
Bandwidth
system, –3 dB, starting at DC
> 250 MHz (meas.)
Input impedance
DC input resistance
system
100 MΩ (meas.)
Input capacitance
system
6.5 pF (meas.)
DC characteristics
Attenuation
system, automatically corrected on base unit
display
100:1
Attenuation error
probe only, with ideal 1 MΩ load impedance
±2 %
Attenuation voltage coefficient
±0.0025 %/V (meas.)
Maximum rated input voltage
Continuous voltage
derated, see figure on page 7
850 V (RMS)
Transient overvoltage
±1200 V
Base unit
Input capacitance
must be compensated by probe’s
LF compensation
10 pF to 50 pF
Input coupling
1 MΩ AC/DC