R&S®RT-Zxx HIGH VOLTAGE AND CURRENT PROBES Specifications Data Sheet | Version 20.
Version 20.00, November 2019 CONTENTS Definitions.............................................................................................................................................................................................. 3 Probe/oscilloscope chart ........................................................................................................................................................................ 4 R&S®RT-ZH03 high voltage passive probe ..................................
Version 20.00, November 2019 Definitions General Product data applies under the following conditions: Three hours storage at ambient temperature followed by 30 minutes warm-up operation Specified environmental conditions met Recommended calibration interval adhered to Specifications with limits Represent warranted product performance by means of a range of values for the specified parameter.
Version 20.
Version 20.
Version 20.00, November 2019 R&S®RT-ZH03 high voltage passive probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 20.
Version 20.00, November 2019 General data Temperature Temperature loading Climatic loading Altitude Safety RoHS Mechanical data Dimensions Weight Probe interface Connector 8 operating temperature range operation diameter of probe tip cable length probe only Rohde & Schwarz R&S®RT-Zxx High Voltage and Current Probes 0 °C to +40 °C 80 % relative humidity without condensation up to 2000 m in line with Low Voltage Directive 2006/95/EC, IEC/EN 61010-31 (pollution degree 2) in line with EN50581 approx.
Version 20.00, November 2019 R&S®RT-ZH10/-ZH11 high voltage passive probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 20.
Version 20.
Version 20.00, November 2019 R&S®RT-ZI10(C)/-ZI11 isolated probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. The R&S RT-ZI10/-ZI11 must be used only with insulated oscilloscopes provided with touch-protected inputs. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 20.00, November 2019 R&S®RT-ZI10C Step response Rise time Frequency response Bandwidth Input impedance DC input resistance Input capacitance DC characteristics Attenuation Base unit Use with Input capacitance Input coupling system, 10 % to 90 % 700 ps (meas.) system, –3 dB, starting at DC > 500 MHz (meas.) system system 10 MΩ ± 1 % (meas.) 11 pF (meas.
Version 20.
Version 20.
Version 20.00, November 2019 R&S®RT-ZD002/-ZD003 high voltage differential probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details. R&S®RT-ZD002 Step response Rise time Frequency response Bandwidth Common mode rejection Input impedance DC input resistance Input capacitance DC characteristics Attenuation Attenuation error Max.
Version 20.
Version 20.
Version 20.00, November 2019 R&S®RT-ZD01 high voltage differential probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details. Attenuation setting Step response Rise time Frequency response Bandwidth Common mode rejection Input impedance DC input resistance Input capacitance DC characteristics Attenuation error Zero error Max.
Version 20.00, November 2019 Maximum rated input voltage Continuous voltage derated, see figure, each signal socket to ground 1000 V (RMS), CAT III Base unit Input coupling 1 MΩ AC/DC Sine-wave RMS voltage [V] 10000 1000 100 10 100E3 1E6 10E6 100E6 Frequency [Hz] Maximum rated sine-wave root mean square voltage versus frequency.
Version 20.
Version 20.00, November 2019 R&S®RT-ZD02/-ZD08 high voltage differential probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details. Step response Rise time Frequency response Bandwidth Common mode rejection Input impedance DC input resistance Input capacitance DC characteristics Max.
Version 20.
Version 20.00, November 2019 R&S®RT-ZHD07 high voltage differential probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 20.00, November 2019 Dynamic range Differential input Offset compensation range Offset compensation error Operating voltage window Noise voltage between signal sockets in both attenuations offset compensation setting = 0 V offset compensation setting ≠ 0 V each signal socket to ground referenced to probe input ±75 V ±750 V ±1000 V no additional error ±0.2 % of setting ±40 mV (meas.) ±750 V 12 mV (RMS) 40 mV (RMS) (meas.) (meas.
Version 20.00, November 2019 Maximum rated sine-wave root mean square voltage versus frequency; each signal socket to ground R&S®ProbeMeter Specifications for measurement error apply only when offset compensation setting is 0 V. The R&S ®ProbeMeter can be used to measure differential and common mode voltages.
Version 20.00, November 2019 General data Temperature Temperature loading operating temperature range storage temperature range Climatic loading Altitude Mechanical resistance Vibration operation transport sinusoidal random Shock EMC Calibration interval Safety RoHS 0 °C to +50 °C −40 °C to +70 °C +25 °C/+40 °C cyclic at 95 % relative humidity without condensation, in line with IEC 60068-2-30 up to 3000 m up to 4600 m 5 Hz to 150 Hz, max. 2 g at 55 Hz, 0.
Version 20.00, November 2019 Mechanical data Dimensions Weight Probe interface Connector Input sockets 28 probe amplifier box, without protector (W × H × L) cable length overall length probe amplifier only Rohde & Schwarz R&S®RT-Zxx High Voltage and Current Probes approx. 55 mm × 26 mm × 165 mm (2.17 in × 1.02 in × 6.5 in) approx. 1.3 m (52 in) approx. 1.6 m (63 in) approx. 360 g (0.
Version 20.00, November 2019 R&S®RT-ZHD15/-ZHD16 high voltage differential probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 20.00, November 2019 Zero error Temperature drift, zero error Dynamic range Differential input Offset compensation range Offset compensation error Operating voltage window Noise voltage Maximum rated input voltage Continuous voltage Transient voltage Base unit Input coupling 30 after applying digital correction factors, referenced to probe input +15 °C to +35°C 0 °C to +50°C referenced to probe input referenced to probe output ±25 mV ±50 mV ±1.5 mV/°C ±30 µV/°C ±65 mV ±95 mV ±1.75 mV/°C ±3.
Version 20.00, November 2019 Maximum rated sine-wave root mean square voltage versus frequency; each signal socket to ground R&S®ProbeMeter Specifications for measurement error apply only when offset compensation setting is 0 V. The R&S ®ProbeMeter can be used to measure differential and common mode voltages.
Version 20.00, November 2019 50/60 Hz rejection Integration time General data See page 27.
Version 20.00, November 2019 R&S®RT-ZHD60 high voltage differential probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 20.00, November 2019 Dynamic range Differential input Offset compensation range Offset compensation error Operating voltage window Noise voltage between signal sockets in both attenuations offset compensation setting = 0 V offset compensation setting ≠ 0 V each signal socket to ground referenced to probe input ±600 V ±6000 V ±2000 V no additional error ±0.2 % of setting ±100 mV (meas.) ±6000 V 70 mV (RMS) 280 mV (RMS) (meas.) (meas.
Version 20.00, November 2019 R&S®ProbeMeter Specifications for measurement error apply only when offset compensation setting is 0 V. The R&S ®ProbeMeter can be used to measure differential and common mode voltages. Measurement error, differential mode and common mode Temperature drift Common mode rejection, for differential measurement 50/60 Hz rejection Integration time +15 °C to +35 °C ≤ 1000 V > 1000 V 0 °C to +50 °C ≤ 1000 V > 1000 V +15 °C to +35 °C 0 °C to +50 °C ±0.12 % of reading ±0.1 V ±0.
Version 20.00, November 2019 R&S®RT-ZC02/-ZC03 current probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details. Sensitivity setting Step response Rise time Frequency response Bandwidth DC characteristics Dynamic range Sensitivity error Temperature drift, sensitivity Zero error R&S®RT-ZC02 0.01 V/A 10 % to 90 % 20 µs (meas.
Version 20.00, November 2019 R&S®RT-ZC03 Step response Rise time Frequency response Bandwidth DC characteristics Dynamic range Sensitivity Sensitivity error Temperature drift, sensitivity Zero error 10 % to 90 % 1 µs (meas.) –0.5 dB, starting at DC 100 kHz (meas.) derated, see figures on page 39 ±30 A 0.1 V/A ±1 % (meas.) ±0.01 %/°C (meas.) ±2 mA (meas.
Version 20.
Version 20.
Version 20.00, November 2019 R&S®RT-ZC05B/-ZC10(B)/-ZC15B/-ZC20(B)/-ZC30 current probes All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 1 MΩ. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 20.00, November 2019 Step response Rise time Frequency response Bandwidth Propagation delay Input impedance DC characteristics Sensitivity Sensitivity error Zero error AC characteristics AC sensitivity error (sinusoidal, 45 Hz to 66 Hz) Measurement due to external magnetic fields Maximum rated input Maximum continuous current Maximum transient current Other Noise R&S®RT-ZC15B R&S®RT-ZC20(B) 10 % to 90 % 7 ns (meas.) 3.5 ns (meas.) –3 dB, starting at DC 50 MHz (meas.) 16.5 ns (meas.
Version 20.00, November 2019 R&S®RT-ZC30 Step response Rise time Frequency response Bandwidth Input impedance DC characteristics Sensitivity Sensitivity error Zero error AC characteristics AC sensitivity error (sinusoidal, 45 Hz to 66 Hz) Measurement due to external magnetic fields Maximum rated input Maximum continuous current Maximum transient current Other Noise 42 10 % to 90 % 2.9 ns (meas.) –3 dB, starting at DC 120 MHz (meas.
Version 20.00, November 2019 General data R&S®RT-ZC05B/ R&S®RT-ZC10(B) Temperature Temperature loading Climatic loading Altitude Safety operating temperature range storage temperature range operation RoHS EMC Calibration interval Mechanical data Dimensions Weight Probe interface Connector Supply voltage max. conductor diameter cable length, probe cable length, power supply of R&S®RT-ZCxx probe head (W × H × L, approx.
Version 20.00, November 2019 Maximum rated sine-wave root mean square input current versus frequency Input impedance (meas.) Maximum rated sine-wave root mean square input current versus frequency Input impedance (meas.
Version 20.00, November 2019 R&S®RT-ZA13 probe power supply Electrical data Number of channels Output voltage Maximum output current Power requirements Maximum rated input power sum total of all channels 4 ±12 V ±0.5 V 2.5 A 100 V to 240 V, 50/60 Hz 170 W General data Safety RoHS EMC Mechanical data Dimensions Weight Connector in line with EN 61010-1 in line with EN50581 in line with EN 61326-1 (class B equipment), EN 61000-3-2, EN 61000-3-3 W×H×L approx. 80 mm × 119 mm × 200 mm (3.1 in × 4.7 in × 7.
Version 20.00, November 2019 Ordering information Designation High voltage passive probes 250 MHz high voltage probe, passive, 100:1, 100 MΩ, 6.5 pF, 850 V (RMS) Incl. adjustment tool; coding clips (set) 2 × 4 colors; signal pin (2); sprung hook 5 mm; ground lead 14 cm; insulating cap; protective cap; operating manual 400 MHz high voltage probe, passive, 100:1, 50 MΩ, 7.5 pF, 1 kV (RMS) Incl. adjustment tool; BNC adapter 5.0-L; coding rings (set) 3 × 4 colors; flexible adapter 5.
Version 20.00, November 2019 Designation Differential probes 25 MHz differential probe, ±700 V, 1 kV (RMS) CAT III, BNC Incl. sprung hook 4 mm (red, black); safety alligator clip 4 mm (red, black); USB power cord; trimming tool; operating manual 25 MHz differential probe, ±1.4 kV, 1 kV (RMS) CAT III, BNC Incl. sprung hook 4 mm (red, black); safety alligator clip 4 mm (red, black); USB power cord; trimming tool; operating manual 100 MHz differential probe, ±1.4 kV, 1 kV (RMS) CAT III, BNC Incl.
Version 20.00, November 2019 Designation Current probes 20 kHz current probe, AC/DC, 0.01/0.001 V/A, 1000 A, 300 V (RMS) CAT III, BNC Incl. operating manual 100 kHz current probe, AC/DC, 0.1 V/A, 30 A, 300 V (RMS) CAT III, BNC Incl. operating manual 10 MHz current probe, AC/DC, 0.01 V/A, 150 A (RMS), BNC Incl. carrying case; operating manual 100 MHz current probe, AC/DC, 0.1 V/A, 30 A (RMS), BNC Incl. carrying case; operating manual 120 MHz current probe, AC/DC, 1 V/A, 5 A (RMS), BNC Incl.
Version 20.00, November 2019 Designation Accessories and sets Mini clips, contains: mini clip (10) Micro clips, contains: micro clip (4) Lead set, contains: lead 6 cm (2.4 in) (5); lead 15 cm (5.
Version 20.
Version 20.00, November 2019 Extended warranty with a term of one to four years (WE1 to WE4) Repairs carried out during the contract term are free of charge 2. Necessary calibration and adjustments carried out during repairs are also covered. Extended warranty with calibration (CW1 to CW4) Enhance your extended warranty by adding calibration coverage at a package price. This package ensures that your Rohde & Schwarz product is regularly calibrated, inspected and maintained during the term of the contract.
Rohde & Schwarz GmbH & Co. KG www.rohde-schwarz.com Worldwide Local und personalized Customized and flexible Uncompromising quality Long-term dependability Regional contact ►► Europe, Africa, Middle East | +49 89 4129 123 45 customersupport@rohde-schwarz.com ►► North America | 1 888 TEST RSA (1 888 837 87 72) customer.support@rsa.rohde-schwarz.com ►► Latin America | +1 410 910 79 88 customersupport.la@rohde-schwarz.com ►► Asia Pacific | +65 65 13 04 88 customersupport.asia@rohde-schwarz.