Table 1-1: Key characteristics
Parameter R&S RT-ZHD07 R&S RT-
ZHD15/16
R&S RT-ZHD60
Bandwidth DC - 200 MHz DC - 100 MHz
(RT-ZHD15)
DC - 200 MHz
(RT-ZHD16)
DC - 100 MHz
Dynamic range
(differential input)
Attenuation high: ±750 V ±1500 V ±6000 V
Attenuation low: ±75 V ±150 V ±600 V
Operating voltage window
(each pin to GND)
±750 V ±1500 V ±6000 V
Differential Offset ±1000 V ±2000 V ±2000 V
Diff. input resistance 5 MΩ 10 MΩ 40 MΩ
Diff. input capacitance 2.5 pF 2 pF 2 pF
R&S ProbeMeter, measurement error < 0.1 % < 0.1 % < 0.12 %
Maximum input voltage to earth (each terminal) 300 V CAT III
600 V CAT II
600 V RMS /
4500 V
PK
1000 V CAT III
1000 V RMS/
6800 V
PK
1000 V CAT III
1750 V RMS/
6800 V
PK
1.1.2 Key Features
Micro button
The micro button is located at the probe head. The micro button can be configured via
the base unit. By pressing it, you can remotely control different functions on the base
unit.
For details, see Chapter 2.6, "Micro Button", on page 25.
R&S ProbeMeter
The R&S ProbeMeter measures the DC voltage of the input signal directly at the probe
tip. It provides a continuous high-precision DC voltage measurement that is independ-
ent of the settings of the oscilloscope and in parallel to the time domain measurement.
If activated on the base unit, the measurement value is displayed on the screen of the
Rohde & Schwarz oscilloscope.
For details, see Chapter 2.8, "R&S ProbeMeter", on page 27.
The R&S ProbeMeter simultaneously records both differential and common mode DC
voltages.