User Manual

Product Description
R&S
®
RT-ZHD07/15/16/60
14User Manual 1800.2588.02 ─ 01
(1) Rohde & Schwarz probe interface with 7 mm
(276 mil) coaxial connector and 6 pogo pins
(2) Release knob
Risk of interface damage
Connect the probe only to a base unit or an adapter that has a Rohde & Schwarz inter-
face. Do not connect the probe to a commercially available BNC jack, as this can
cause irreparable damage to the interface. From the outside, the 7 mm coaxial connec-
tor looks like a standard BNC connector. However, it is constructed differently and is
capable of a significantly higher frequency limit.
Shock hazard caused by high voltages
The attached input leads have a jacket wear indicator. If the input lead's jacket is
excessively worn, a different jacket color becomes visible. If you see this color indica-
tor, do not use the probe lead.
If other input leads were used, make sure that they fulfil the safety requirements for
your measurement. Do not use 4 mm banana plugs without protection against contact.
Influence of the accessory to the CAT rating
The CAT rating goes along with the minor rating of the accessory used for a specific
measurement. E.g. the usage of the spade terminals reduces the CAT of the whole
measurement set to CAT II. This is independent from the used R&S RT-ZHD.