Putting into Operation
R&S
®
RT-ZHD07/15/16/60
23User Manual 1800.2588.02 ─ 01
The oscilloscope settings for attenuation and offset are automatically adjusted.
As soon as the probe is connected to the oscilloscope and the settings are adjusted,
the waveform is shown for the channel to which the probe is connected.
1. On the "Vertical" menu, select "Probe Setup".
2. On the left side of the "Setup" tab, select the channel to which the probe is connec-
ted.
The complete probe information is shown on the "Probe Attributes" tab.
2.4 Connecting the Probe to the DUT
Shock hazard caused by high voltages
Before connecting the probe to the test circuit, make sure that probe is connected to
the measuring instrument and the instrument is properly grounded.
Ensure a stable connection between the DUT and the probe. Switch off the test circuit
while connecting and disconnecting the probe leads.
The finger guard provides protection. Keep your fingers behind the finger guard.
1. Connect the clips to the input leads (4 mm safety banana plugs).
2. Do not touch the unsafe area during a measurement.
Connecting the Probe to the DUT