Putting into Operation
R&S
®
RT-ZHD07/15/16/60
25User Manual 1800.2588.02 ─ 01
Figure 2-4: Leads, untwisted
Figure 2-5: Leads, twisted
2.6 Micro Button
The micro button provides easy and quick access to important functions of the
Rohde & Schwarz oscilloscope. After a function has been assigned, pressing the micro
button remotely controls this specific function on the base unit. For example, "Continu-
ous Run" and "AutoZero" are commonly assigned to the micro button.
To assign a function to the micro button with the Rohde & Schwarz oscilloscope:
"Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action"
For more details on the available functions and settings for the micro button, refer to
the "User Manual" of the Rohde & Schwarz oscilloscope.
2.7 Offset Compensation
The offset compensation function can compensate a DC component of the input signal
between the positive and negative input, even in front of the active amplifier in the
probe control box. This is particularly helpful if single-ended signals are measured with
a differential probe, for example, with the negative input socket connected to ground.
These signals often have a superimposed DC component, which can be compensated
using the offset compensation on the probe.
Offset Compensation