Table 1-2: Attenuation and maximum peak voltages of R&S
RT-ZHD
Probe type Attenuation high (att. high) Attenuation low (att. low)
Att. Max. Voltage Att. Max. Voltage
RT-ZHD07 250 : 1 750 V
pk
25 : 1 75 V
pk
RT-ZHD15/16 500 : 1 1500 V
pk
50 : 1 150 V
pk
RT-ZHD60 1000 : 1 6000 V
pk
100 : 1 600 V
pk
The setting with lower attenuation is characterized by less noise.
The selected attenuation does not influence the allowed working voltage, measure-
ment category, and the usable common mode range. Thus, setting the attenuation
does not cause a hazardous situation or a measurement error due to inadmissible
common mode voltages.
Audible overrange on/off
Whenever an overrange condition occurs, the buzzer generates a continuous audible
alarm. This button toggles ON/OFF of the audible signal.
Overrange indicator at the input terminals
The overrange indicator lights red if the voltage of the single ended input signal to GND
exceeds the dynamic input range limit of the probe. In this case, the signal on the
probe output may not accurately represent the signal on the probe input.
Overrange indicator at the output terminal
The overrange indicator lights red if the voltage of the differential signal exceeds the
dynamic range limit at the output of the probe. In this case, the signal on the probe out-
put may not accurately represent the signal on the probe input.
Micro Button
The micro button provides easy and quick access to important functions of the
Rohde & Schwarz oscilloscope. After a function has been assigned, pressing the micro
button remotely controls this specific function on the base unit. For example, "Continu-
ous Run" and "AutoZero" are commonly assigned to the micro button.
To assign a function to the micro button with the Rohde & Schwarz oscilloscope:
"Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action"
For more details on the available functions and settings for the micro button, refer to
the "User Manual" of the Rohde & Schwarz oscilloscope.
Description of the Probe