Datasheet
Version 12.01, January 2017
6 Rohde & Schwarz R&S
®
RTH1002, R&S
®
RTH1004 Scope Rider
Trigger system
(see also R&S
®
RTH-B1 mixed signal option)
Trigger level
range
±4 div from center of screen
Trigger modes
auto, normal, single
Trigger sources
R&S
®
RTH1004
CH1, CH2, CH3, CH4
R&S
®
RTH1002
CH1, CH2
Hold-off range
time
8 ns to 10 s, fixed and random
events
1 to 1 000 000 000 events
Trigger types
Edge
triggers on specified slope (positive, negative or either) and level
Glitch
triggers on glitches of positive, negative or either polaritiy that are shorter or longer
than specified width
glitch width
200 ps to 5000 s
Width
triggers on positive or negative pulse of specified width; width can be shorter, longer,
inside or outside the interval
pulse width
200 ps to 5000 s
TV/video
triggers on baseband analog progressive and interlaced video signals PAL, NTSC,
SECAM, PAL-M, SDTV and HDTV broadcast standards
(SDTV and HDTV require R&S
®
RTH-K19 option)
trigger events
all fields, odd fields, even fields, all lines,
line number
Pattern
triggers when a logical combination (and, nand, or, nor) of the input channels stays true
for a period of time shorter, longer, inside or outside a specified range
(requires R&S
®
RTH-K19 option)
pattern time
800 ps to 5000 s
State
triggers when a logical combination (and, nand, or, nor) of the input channels stays true
at a slope (positive, negative or either) in one selected channel; state values may be
high (H), low (L) or don’t care (X) (requires R&S
®
RTH-K19 option)
Runt
triggers on pulse of positive, negative or either polarity that crosses one threshold but
fails to cross a second threshold before crossing the first one again; runt pulse width
can be arbitrary, shorter, longer, inside or outside the interval
(requires R&S
®
RTH-K19 option)
runt pulse width
200 ps to 5000 s
Slew rate
triggers when the time required by a signal edge to toggle between user-defined upper
and lower voltage levels is shorter, longer, inside or outside the interval; edge slope
may be positive, negative or either (requires R&S
®
RTH-K19 option)
toggle time
200 ps to 5000 s
Window
triggers when signal enters or exits a specified voltage range; triggers also when signal
stays inside or outside the voltage range for a specified period of time
(requires R&S
®
RTH-K19 option)
window time
200 ps to 5000 s
Data2clock
triggers on setup time and hold time violations between clock and data present on any
two input channels; monitored time interval may be specified by the user with a step
size of 800 ps in the range from –124 ns to 124 ns around a clock edge
(requires R&S
®
RTH-K19 option)
Serial pattern
triggers on serial data pattern up to 32 bit clocked by one input channel; pattern bits
may be high (H), low (L) or don’t care (X); clock edge slope may be positive, negative
or either (requires R&S
®
RTH-K19 option)
max. data rate
< 250 Mbps
Timeout
triggers when signal stays high, low or unchanged for a specified period of time
(requires R&S
®
RTH-K19 option)
timeout
200 ps to 5000 s
Interval
triggers when time between two consecutive edges of same slope (positive or
negative) is shorter, longer inside or outside a specified range
(requires R&S
®
RTH-K19 option)
interval time
200 ps to 5000 s
Protocol