Datasheet

Version 12.01, January 2017
Rohde & Schwarz R&S
®
RTH1002, R&S
®
RTH1004 Scope Rider 7
Waveform measurements
(see also R&S
®
RTH-B1 mixed signal option)
Automatic measurements
total number of active measurements
4
sources
R&S
®
RTH1004
CH1, CH2, CH3, CH4, math, reference
R&S
®
RTH1002
CH1, CH2, math, reference
time based measurements
period, frequency, rise time, fall time,
positive pulse width, negative pulse width,
positive duty cycle, negative duty cycle,
delay, phase
amplitude based measurements
mean value, RMS value, crest factor,
standard deviation, minimum, maximum,
peak-to-peak, base level, top level,
amplitude, overshoot, preshoot, AC, DC,
AC+DC
count based measurements
count positive pulses, count negative
pulses, count rising edges, count falling
edges
power based measurements
active power, apparent power, reactive
power, power factor
Cursor measurements
sources
analog channels, math and reference
waveforms
vertical
2 cursors showing time, time difference
and inverse time difference (frequency)
horizontal
2 cursors showing voltage and voltage
difference
tracking
vertical cursor additionally showing
voltage and voltage difference of selected
waveform
measure
defines gate for automatic measurements
Mask testing
Sources
R&S
®
RTH1004
CH1, CH2, CH3, CH4, math
R&S
®
RTH1002
CH1, CH2, math
Mask definition
tolerance tube based on analog input
waveform or math waveform
Number of simultaneous mask tests
up to 5
Actions on violation
none, beep, stop
History behavior
requires R&S
®
RTH-K15 option
store all
Waveform maths
Number of math waveforms
1
Functions
addition, subtraction, multiplication,
square, absolute value, inverse
Sources
R&S
®
RTH1004
CH1, CH2, CH3, CH4
R&S
®
RTH1002
CH1, CH2
FFT
Number of simultaneous spectra
1 (always on selected channel)
Sources
R&S
®
RTH1004
CH1, CH2, CH3, CH4
R&S
®
RTH1002
CH1, CH2
FFT sizes
64 ksample or 8 ksample
Sampling frequency
equal to sampling rate of current
horizontal scale for FFT size of 64
ksample;
equal to 1/8
th
of sampling rate of current
horizontal scale for FFT size of 8 ksample
Window types
rectangular, flat-top, Hamming, Hann,
Blackman
Channel bandwidth
same as bandwidth limits in vertical
system;
additionally accessible over FFT menu
X-axis scaling
linear or logarithmic