User Manual

Waveform Analysis
R&S
®
Scope Rider RTH
75User Manual 1326.1578.02 ─ 08
Note: The zoom overview also considers the horizontal position of the trigger point. If
the horizontal position is 0 and thus the trigger point is not in the middle, the zoom
area in the overview is also shifted, even if the zoom position is 0.
Remote command:
ZOOM:POSition on page 287
3.2 Automatic Measurements
You can perform up to four different measurements simultaneously.
3.2.1 Performing Automatic Measurements
To start and stop the last configured measurements
Press the MEAS key.
To configure automatic measurements in the Meas menu
1. Press and hold the MEAS key until the "Meas" menu opens.
2. Select the number of the measurement that you want to configure.
3. If the measurement is disabled, enable "State".
4. Select the "Type".
The selection list shows all available measurement types.
5. Select the "Source".
The selection list shows all all active sources that are allowed for the selected mea-
surement type.
6. Some measurement types require additional settings. Scroll down the menu and
adjust the additional settings if necessary.
3.2.2 Measurement Results
The measurement results are shown on the left-top side of the screen.
Figure 3-3: Results of 4 active measurements
If a result cannot be determined, "---" is displayed. Adjust the horizontal and vertical
settings if the instrument cannot measure.
Automatic Measurements