User Manual

Waveform Analysis
R&S
®
Scope Rider RTH
81User Manual 1326.1578.02 ─ 08
Measurement
Selects the measurement to be configured in the menu. You can perform up to four dif-
ferent measurements simultaneously.
State
Enables or disables the selected measurement.
Remote command:
MEASurement<m>:ENABle on page 287
Type
Selects the measurement type. For a detailed description, see Chapter 3.2.3, "Mea-
surement Types", on page 76.
Remote command:
MEASurement<m>:TYPE on page 288
Source / Source 2
Defines the waveform to be measured. For delay, phase, and power measurements, 2
sources are required.
The sources can be any active input signal, math or reference waveform. Available
source waveforms depend on the measurement type, see Chapter 3.2.3, "Measure-
ment Types", on page 76.
Remote command:
MEASurement<m>:SOURce on page 288
All off
Disables all active measurements.
Remote command:
MEASurement<m>:AOFF on page 289
Slope
Sets the slope for the delay measurement type.
"Positive"
Delay between the first rising edge of each source waveform.
"Negative"
Delay between the first falling edge of each source waveform.
"Either"
Delay between the first edge of each source waveform, no matter if it
is rising or falling.
Remote command:
MEASurement<m>:DELay:SLOPe on page 289
3.3 Cursor Measurements
The cursor measurement determines the results at the current cursor positions, or per-
forms gated automatic measurements between the cursor lines. The cursors can be
positioned manually at fixed positions, or they can follow the waveform.
Cursor Measurements