User Manual

Waveform Analysis
R&S
®
Scope Rider RTH
86User Manual 1326.1578.02 ─ 08
To activate or deactivate the last configured math waveform, shortly press the
MATH key.
To adjust vertical scale and position of the math waveform, use the RANGE and
POS keys.
You can analyize math waveforms in the same way as channel waveforms: use zoom,
perform automatic and cursor measurements, save as reference waveform, and per-
form mask tests.
Description of settings
State
Activates the waveform and displays it.
Remote command:
REFCurve:STATe on page 296
CALCulate:MATH:STATe on page 294
Source 1 / Source 2
Sets the source(s) for the defined mathematic operation.
Operation
Select an operation to calculate the math waveform.
"S1 + S2"
Addition: Adds the values of "Source 1" and ""Source 2"".
"S1 - S2"
Subtraction: Subtracts the values of "Source 2" from the values of
"Source 1".
Mathematics