User Manual

Mask Testing
R&S
®
Scope Rider RTH
99User Manual 1326.1578.02 ─ 08
Figure 4-1: Mask-mode-view
1 = Selected channel
2 = Number and percentage share of passed acquisitions
3 = Number and percentage share of failed acquisitions
4 = Number of tested acquisitions
5 = Test duration
Remote commands for mask test results:
MASK:CHANnel<m>:RESult:PASS[:COUNt]? on page 303
MASK:CHANnel<m>:RESult:PASS:PERCentage? on page 303
MASK:CHANnel<m>:RESult:FAIL[:COUNt]? on page 302
MASK:CHANnel<m>:RESult:FAIL:PERCentage? on page 302
MASK:CHANnel<m>:RESult:TOTL[:COUNt]? on page 303
MASK:ELAPsedtime:TOTal? on page 303
MASK:ELAPsedtime[:SECS]? on page 304
MASK[:TESTstate]? on page 302
4.3 Running Mask Tests
1. Before you can perform mask tests, adjust the waveforms by editing the vertical,
horizontal, trigger and acquisition settings.
For details see:
Chapter 2.2, "Vertical Setup", on page 38
Chapter 2.3, "Horizontal Setup", on page 42
Chapter 2.5, "Trigger", on page 46
Chapter 2.4, "Acquisition Control", on page 43
Note: The acquisition mode "Envelope" is not available while running mask tests.
Running Mask Tests