User Manual
Spectrum Analysis
R&S
®
Scope Rider RTH
118User Manual 1326.1578.02 ─ 08
You can configure which peaks the instrument will find within the spectrum results.
Peaks are indicated by markers in the power vs frequency display. The frequency and
power level measured at the peak is also indicated in the diagram.
Overlapping markers
If two detected peaks are very close to each other and the markers in the display over-
lap, the marker with the lower amplitude is indicated in a darker color and put slightly in
the background, while the one with the higher amplitude is displayed normally. Using
the remote commands, both values can be retrieved.
Remote commands for results:
SPECtrum:MARKer:RESult<m>:FREQuency[:VALue]? on page 316
SPECtrum:MARKer:RESult<m>:FREQuency:DELTa? on page 316
SPECtrum:MARKer:RESult<m>:LEVel[:VALue]? on page 317
SPECtrum:MARKer:RESult<m>:LEVel:DELTa? on page 317
SPECtrum:MARKer:RMARker:FREQuency? on page 317
SPECtrum:MARKer:RMARker:VALue? on page 318
State............................................................................................................................119
Number of Markers..................................................................................................... 119
Threshold.................................................................................................................... 119
Peak excursion............................................................................................................119
Distance...................................................................................................................... 120
Trace Source...............................................................................................................120
Marker Value...............................................................................................................120
Spectrum Mode (Option R&S