User Manual
Spectrum Analysis
R&S
®
Scope Rider RTH
131User Manual 1326.1578.02 ─ 08
8. To calculate the average results over several measurements, select "Average" and
define the number of results to consider.
9. To clear the result data used for statistical evaluation and start a new measure-
ment, select "Clear Results".
To save the results to a file
1. Press the FILE key.
2. Select "Harmonics".
3. Check the "Harmonic Directory" and the "Filename Base". Adjust if necessary.
4. Select "Save".
For details on the format of the result file, see Chapter 5.3.6.2, "Harmonic Results
Export File Format", on page 135.
5.3.4 Harmonic Mode Settings
Access: "Harmonic" menu
CH <N>....................................................................................................................... 131
Scale........................................................................................................................... 131
THD Type....................................................................................................................131
Statistic........................................................................................................................132
Clear Results...............................................................................................................132
Fundamental Frequency, Value.................................................................................. 132
Limits, Loaded File...................................................................................................... 132
Open........................................................................................................................... 133
Display Harmonics, Max No Harmonics......................................................................133
Average.......................................................................................................................133
CH <N>
Selects the channel for which the captured data is analyzed in "Harmonic" mode. Har-
monic measurements can be performed on all channels simultaneously. The results
are indicated by different colored bars and numbers, corresponding to the channel but-
ton color.
Remote command:
CHANnel<m>:STATe on page 252
Scale
Switches the scale for the measured harmonic levels between logarithmic (dB) and lin-
ear (percent). The values are relative to the level measured for the fundamental fre-
quency.
Remote command:
HARMonic:SCALe on page 322
THD Type
Switches between different calculation methods for the total harmonic distortion.
Harmonics Measurement (Option R&S RTH-K34)