User Manual

Spectrum Analysis
R&S
®
Scope Rider RTH
135User Manual 1326.1578.02 ─ 08
ted, displayed, and exported. A limit check is performed only for those harmonics for
which a limit value is defined.
Sample file
A sample limit file is provided on the instrument:
C:\Users\<user_name>\Rohde-Schwarz\RTH\Harmonic\LimitExample.csv
The file format is a comma-separated list (.csv). At least the following contents must
be included:
Header:
<Device>,Version,<file_version>
Harmonics,Limit[%]
Values:
<Order of harmonic>, <limit value in % of fundamental>
Comments introduced by "#" are allowed.
If a limit is not provided for a harmonic, the harmonic is displayed and its results
are included in the export, but not in the limit check.
The fundamental is always displayed and need not be included in the list.
Example:
RTH,Version,1.0
Harmonics,Limit[%]
2,10.5
4,0.004
20,3.7
10
13,0.01
3,0.01
5.3.6.2 Harmonic Results Export File Format
The harmonic results export file format is similar to a waveform export file; it is also
stored in CSV format. You can convert the comma-separated text to columns (see
Chapter 11.4.3.3, "Converting CSV to Excel Files", on page 229).
The harmonics results file contains some general header data and then a set of result
data for each harmonic. A maximum of 64 results can be provided, but the actual num-
ber of harmonics may be restricted (see "Limits, Loaded File" on page 132). Each set
of data contains the results for a specific harmonic in each input channel.
Table 5-5: Header data (converted to table)
Harmonic Results
[Description - not included in file]
Model RTH1004
Harmonics Measurement (Option R&S RTH-K34)