User Manual
Multimeter Measurements
R&S
®
Scope Rider RTH
138User Manual 1326.1578.02 ─ 08
6.1.2 Display and Control
The multimeter display shows detailed results and setup icons.
Figure 6-1: Display of the digital multimeter
1 = Measurement state. "Manual" or "Auto": running measurement with manual or automatic range. "Hold":
stopped measurement.
2 = Measurement range
3 = Test lead connection including polarity
4 = Minimum value and timestamp
5 = Average value
6 = Maximum value and timestamp
7 = Warning sign, shows up if the measured value is higher than 30 V
8 = Bargraph displaying the measured value
9 = Measurement type
10 = Activates or deactivates relative measurement
11 = Restarts the measurement and resets all values
In meter mode, the keys behave slightly different from scope mode:
●
The AUTOSET key toggles the range mode: autoranging and manual range set-
ting.
●
The vertical RANGE and POS keys adjust the measurement range.
●
The MEAS key opens the "Meter" menu.
●
The following keys work as usual: FILE,
, , PRESET, MODE, BACK.
●
The CH keys switch back to the scope mode.
●
All other keys do not work.
6.1.3 Running Multimeter Measurements
1. Connect the test leads to the meter inputs.
Digital Multimeter (R&S RTH1002)