User Manual
Multimeter Measurements
R&S
®
Scope Rider RTH
143User Manual 1326.1578.02 ─ 08
Figure 6-4: Four-voltmeter-view
1 = Channel measured by the voltmeter
2 = Bargraph displaying the measured value
3 = Measurement ranges of active channels
In meter mode, the keys behave different from scope mode:
●
The vertical RANGE and POS keys both adjust the measurement range of the
selected channel.
●
The MEAS key opens the "Meter" menu.
●
The following keys work as usual: FILE,
, , PRESET, MODE, BACK.
●
All other keys do not work.
Displaying voltmeters
► To switch to one-voltmeter-view, use one of the following methods:
● Tap the meter number as shown in Figure 6-4.
● Turn the wheel to select a voltmeter, and press the wheel button to open it.
► To view all 4 voltmeters, use one of the following methods:
● Tap the highlighted number of the opened voltmeter as shown in Figure 6-3.
● Press BACK.
6.2.3 Running Voltmeter Measurements
1. Set up the channels for voltmeter measurements.
a) Select the "Vertical" menu.
Voltmeter (R&S RTH1004)