User Manual
Data Logging
R&S
®
Scope Rider RTH
148User Manual 1326.1578.02 ─ 08
Figure 7-1: Logger display, view of all logged scope measurements
1 = Logger channel, each logger channel records one measurement
2 = Latest values of the logged measurements (depends on logging type and measurement type)
3 = Time basis
4 = In run mode (recording): →Slot <x> is the selected slot. In stop mode: Slot <x> is the active slot where
data has been stored to or loaded from
5 = Recording time
6 = Start time
7 = Upper and lower scaling value of each measurement
8 = Time stamp of the left side of the display
Displaying logged measurements
1. To switch to one-log-view, use one of the following methods:
● Tap on one of the logged measurements as shown in Figure 7-1.
● Press the assigned CH key.
2. To view all logs, use one of the following methods:
● Tap the highlighted number of the open logger channel.
● Press the CH key of the open logger channel.
7.3 Logging Data
You can log up to four different scope measurements or meter measurements simulta-
neously.
Before you start the logging, set up and activate the measurements that you want to
log. See:
●
Chapter 3.2, "Automatic Measurements", on page 75
●
Chapter 6, "Multimeter Measurements", on page 137
Logging Data