User Manual
General Instrument Setup
R&S
®
Scope Rider RTH
234User Manual 1326.1578.02 ─ 08
12 General Instrument Setup
The general instrument settings are available in all operating modes.
12.1 Resetting the Instrument
Reset is helpful if the instrument is in undefined condition and cannot be operated.
►
To switch off the instrument and reset the hardware, press the
power key for 5
seconds.
► To restore the factory settings, hold down the PRESET key during the boot proc-
ess.
12.2 Disabling the Touchscreen
If you use the instrument in an environment with immissions considerably higher than
specified, the immissions can affect the touch sensitivity of the screen. In this case,
operate the instrument using the keys and the wheel and disable the touchscreen.
► To disable the touchscreen, press the SHIFT key for 2 seconds.
12.3 Selfalignment
Access: or "Setup" menu > "Selfalignment"
The self-alignment aligns the data from several input channels vertically and horizon-
tally to synchronize the timebases, amplitudes and positions.
The R&S RTH1002 provides an additional, separate self-alignment of the meter inputs.
Recommendation on performing the self-alignment:
●
When putting the instrument into operation for the first time
●
After a firmware update
●
Once a week
●
When major temperature changes occur (> 5°)
Selfalignment