User Manual

General Instrument Setup
R&S
®
Scope Rider RTH
240User Manual 1326.1578.02 ─ 08
"Time"
Data points remain on the screen for the duration defined with Persis-
tence Time.
"Infinite"
Data points remain on the screen infinitely until persistence is set to
"Off".
Remote command:
DISPlay:PERSistence[:TYPE] on page 437
Persistence Time
User-defined persistence time if "Persistence Type" is "Time". Each new data point
remains on the screen for the duration defined here.
Remote command:
DISPlay:PERSistence:TIME on page 438
Grid Mode
Defines the grid display. A grid helps you identify the position of specific data points.
"Off"
No grid is shown.
"Dots"
Dots mark the intersections of the gridlines.
"Lines"
Displays the grid as horizontal and vertical lines.
Brightness
Changes the brightness of the touchscreen.
Show Probe Settings
Defines how long the probe settings of active channels are shown when you press a
channel key. The settings are shown on the top of the display.
The probe settings are always shown as long as the "Vertical" menu is open.
Display Off After
Defines when the display switches off if it is not used. Switching off the display saves
energy and prolongs the battery's operating time.
12.7 Options
All options are activated by license keys. No additional installation or hardware change
is required.
Unregistered licenses
Unregistered licenses are not assigned to a particular instrument. The instrument
accepts only registered licenses. If your license is delivered unregistered, use the
online tool R&S License Manager to register the license for your instrument. The regis-
tration of a permanent license is irreversible, so ensure that you register it for the cor-
rect instrument. The address of the tool is https://extranet.rohde-schwarz.com/service.
The "Active options" tab provides information on installed software options. Here you
can install new options or deactivate existing options using license keys.
Options