User Manual
Remote Control Commands
R&S
®
Scope Rider RTH
256User Manual 1326.1578.02 ─ 08
Parameters:
<ThresholdType> TTL | ECL | CMOS | CAN | GND | LIN7vsupply | LIN12vsupply |
LIN18vsupply | USER
TTL
1.4 V
ECL
-1.3 V
CMOS
2.5 V
GND
0 V (for CAN channels, requires option R&S RTH-K3)
CAN
2 V (for CAN channels, requires option R&S RTH-K3)
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V (for LIN channels, requires option R&S RTH-
K3)
USER
Set the value with CHANnel<m>:THReshold:USER.
*RST:
TTL
CHANnel<m>:THReshold:USER <ThresholdValue>
Set an individual threshold value if CHANnel<m>:THReshold:TECHnology is set to
USER.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
CHANnel<m>:THReshold:THReshold?
Returns the threshold value.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Return values:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
Waveform Setup