User Manual

Remote Control Commands
R&S
®
Scope Rider RTH
258User Manual 1326.1578.02 ─ 08
Time scale Min./max. horizontal position
1 ns/div to 100 µs/div ±2 s
200 µs/div to 2 s/div Time scale * 20000
5 s/div – 500 s/div 100000 s
TIMebase:REFerence <ReferencePoint>
Defines the time reference point in the diagram.
Parameters:
<ReferencePoint> Position of the reference point in percent of the screen width.
Available values are:
10: on the left side of the screen
50: in the middle of the screen
90: on the right side
*RST: 50
Default unit: %
ACQuire:POINts:ARATe?
Returns the number of recorded analog waveform points per second.
Return values:
<ADCSampleRate> Range: 1.25E+9 to 5E+9
Increment: 1
*RST: 1.25E+9
Default unit: Sa/s
Usage: Query only
14.3.4 Acquisition Control
RUN............................................................................................................................. 258
STOP........................................................................................................................... 259
ACQuire:MODE............................................................................................................. 259
ACQuire:AVERage:COUNt............................................................................................. 259
ACQuire:ARESet:IMMediate........................................................................................... 259
ACQuire:POINts[:VALue]?.............................................................................................. 260
ACQuire:RESolution?.....................................................................................................260
RUN
Starts the continuous acquisition.
Usage:
Event