User Manual
Remote Control Commands
R&S
®
Scope Rider RTH
264User Manual 1326.1578.02 ─ 08
Parameters:
<RandomMinTime>
<RandomMaxTime>
Range: 8E-9 to 10
Increment: 200E-6
*RST: 2E-3
Default unit: s
TRIGger:MNR <MoreNoiseReject>
Enables a hysteresis to avoid unwanted trigger events caused by noise oscillation
around the trigger level.
Parameters:
<MoreNoiseReject> ON | OFF
*RST: OFF
14.3.5.2 Edge Trigger
TRIGger:EDGE:SLOPe <Slope>
Sets the edge direction for the trigger. You can trigger on:
●
Rising edge, that is a positive voltage change
●
Falling edge, that is a negative voltage change
●
Rising and falling edge
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
14.3.5.3 Glitch Trigger
See also: Chapter 2.5.3, "Glitch Trigger", on page 50
TRIGger:GLITch:POLarity...............................................................................................264
TRIGger:GLITch:RANGe................................................................................................ 265
TRIGger:GLITch:WIDTh................................................................................................. 265
TRIGger:GLITch:POLarity <Polarity>
Sets the pulse polarity, that is the direction of the first pulse slope. You can trigger on:
●
Positive going pulses. The width is defined from the rising to the falling edge.
●
Negative going pulses. The width is defined from the falling to the rising edge.
●
Both positive and negative going pulses
Parameters:
<Polarity> POSitive | NEGative | EITHer
*RST: POSitive
Waveform Setup